Operation lamp has been widely used in hospitals and clinics, operation shadowless lamp is simple, easy to use, in order to better play its advantages, we need to know its correct debugging method
One of the debugging of the surgical shadowless lamp - device inspection: mainly to see that all the screws are in place and tightened during the installation process, whether various decorative covers have been covered, or whether there are other devices missing.
The second debugging of the surgical shadowless lamp - circuit inspection: This is a key to the safety inspection of the surgical shadowless lamp. The first is to check whether the shadowless lamp has a short circuit or open circuit in the case of power failure. If not, check whether the power supply of the shadowless lamp is stable after power on. Whether the input voltage of the transformer is stable and meets the requirements of shadowless lamps.
The third debugging of the surgical shadowless lamp - balance arm adjustment: When medical staff adjust the position of the surgical shadowless lamp, they all need the balance arm system to bear the force, so it is necessary to check whether the balance arm can be adjusted to the viewpoint required by the medical staff and whether it can bear the force.
The fourth debugging of surgical shadowless lamp - joint sensitivity: Because the viewpoint of shadowless lamp needs to be adjusted, the sensitivity of the joint is also very important, mainly adjusting the damping screw of the joint. The standard rule is that the tightness of the damping adjustment is the force of advancing or rotating the joint in any direction at 20N or 5Nm.
The fifth debugging of surgical shadowless lamp - illumination depth: because the doctor is likely to need to watch the depth of the patient's trauma during surgery, the surgical shadowless lamp needs to have a good illumination depth, generally a distance of 700-1400mm is better
The sixth debugging of surgical shadowless lamp - illumination and color temperature inspection: This is the more important point of surgical shadowless lamp. Excellent illumination and color temperature help doctors to carefully observe the patient's trauma, distinguish organs, blood, etc., so it is close to the illumination of sunlight and 4400 -4600K color temperature is more suitable.
Post time: Mar-09-2022